IB-100 Series Hyperbaric Chambers
[Standard Series| IB-130 Series| Amagnetic Series]
Steel
manufacture; inner diameters between 1.000 and 1.800 millimetres.
IB-100 Series Hyperbaric Chamber
[Description | Technical
Specifications| Equipment | Special
quipment| Other Specifications ]
![]()
Special manufacturing
IB-1.8
x 4.5 Hyperbaric Chamber 7.5 bar
[Description | Technical
Specifications | Equipment | Optional Equipment | Other
Specifications ]
DESCRIPTION
1.-
This Multiplace Chambers is specially designed for HBO Treatment, diving
training and recovery, research works, equipment controls, leaking test, etc
2.- The
chamber is to be made of steel and will be installed on shore
3.- One
unit
4.-
Model: IB-1.8 x 4.5
5.- Country of origin: Spain
|
|
|||||||||||||||||||
|
6.- TECHNICAL SPECIFICATIONS AT A GLANCE
·
·
Door diameter..... 700 mm ( 800 mm
when bayonet ring dock Adivp-1
NATO in outer door)
Those technical
details are our standard manufacturing characteristics. Inner lock Chamber (inside):
Outer
- lock (inside):
7.- Installation
materials
8.- Installation works A qualified engineer
will supervise all mounting works at sight from compressor, piping, safety systems,
electrical circuits. All workers will
be experienced in that hyperbaric system.
Nobody will be alone without a close supervision of Iberco΄s
manufacturer. 9.- Air supply system a.- Primary Enough to pressurise
the inner lock once and outer lock twice ( are needed for this purpose
100,000 litres air = All air
contained in 10 cylinder 50 liter/200 bar) and will have moreover 18 x 50 x
200 = 180,000 litres in normal conditions (1bar and 20 ΊC) of air to be supplied
from HP reserve cylinder for repeat pressurisations operations , if
necessary. Total offer: 28 cylinder 50 litre/ 360 bar, double head (valve air out superior level
and a lower purge for condensating in each cylinder) b.- Secondary supply. A LP compressor at 13 bar
with a discharge flow up to 3.3 m3/min (30 Kw powered) Ingersoll
Rand type MU-30, will possibility a rate of pressure amount around 4 meter/min in inner lock, enough
for that use and will provide a ventilation flow better than 300 litre/min
(18,000 litre volume tank at 12 bar working pressure) 10.- Pressure relief
system a.
a. Exhaust system should have capacity to
release at the rate of 35 ft sw/min b.
b. Flow meter capable from 150 300 litre/min,
in the inner lock c.
c. Relief valve, one in each lock, with a
shut-off valve before it. 11.- Lighting
12.- Communications -
-
Permanent dynamic
phone and other communications sets are provided for each dock, sound powered
and press to talk dual. -
-
13.- Oxygen supply system -
-
6 oxygen breathing
units with overboard dumping outside the chamber in inner lock. -
-
2 oxygen breathing
units with overboard dumping outside the outer lock. |
|||||||||||||||||||
|
14.-
STANDARD EQUIPMENT Control Panel: Control
and Supervision Boards It is usually attached
to the chamber, with a compact and easy-use design: - Upper frame:
manometers controls and main air
& oxygen supply, depth gauges both inside and outside in each
lock, analysis instruments (CO2, O2), thermometers -Middle frame:
valves for oxygen supply; electrical switchboard control (communications
double systems self-exciting sound powered and press to talk, lights, etc) - Bottom frame:
valves for pressurisation, depressurisation, air renewal, breathing dumping,
flowmeter capable from 150 300 litre per min, etc |
|||||||||||||||||||
|
IB-180x 4.5 model is adapted to specific solutions according
to special requirements: -
-
Check manometer. -
-
Operating control
valves for air and oxigen supply and exhaust - Filtering Systems - High / middle
pressure air supply control boards (valves, high flow reducers, filters) |
|||||||||||||||||||
|
OTHER
TECHNICAL DATA Design code = ASME
VIII-Div. 1, PVHO-01 & Det Norske Veritas, Certification by Cualicontrol
-ACI, SA, (a independent UE qualified Inspection Company authorised by
Spanish Industrial Ministry). Non destructive test: X-ray 100 % Pneumatic Test: (three test). No leakage bigger than 2% in 24 hour
test. Capacity Discharge of safety valves test. Pressurisation Speed test: bigger than 10 msw/min Depressurisation Speed test: bigger than 10 msw/min Functional and working Test at
maximun pressure INSTRUMENTS.
Manometers = WIKA, Class =
0,6; 0 9,8 bar & 0 - 100 mca & 0-250 ft sw VALVES and
OTHER COMPONENTS. Pressurisation / depressurisation valves: = needle valves, brass. 15.- Doors 04 in number doors of
700 mm diameter with rubber seal. Door face should be equipped to adjust with
portable Recompression Chamber. 16.- Medical lock Supply lock main
chamber free diameter....... 300 mm. (medical) type bayonet and double safety
system. 17.-
Berthing system- Stainless steel bench
or berths sould be fitted 18 View
ports - 03 in number
pressure proof view ports (02 for inner locks & 01 for outer lock) should
be provided. 19.- FIRE
PREVENTION -2 Fire extinguisher
with 6 litres desionissed water each one IN INNER LOCK and only one in outer
lock. 20.-
Medical Kit Primary: Flash light Stethoscope Otoscope Sphygmomanometer
(aneroid type) Reflex hammer Tunning fork Pin & brush for
sensory testing Tongue depressors Secondary: Cuffed endotracheal
tubes adapters (7 to 9.5 mm) Syringe and sterile water
cuff inflation (10 cc) Malleable stylet 12
inch length Laryngoscope with
extra batteries and light bulbs Laryngoscope blades Sterile lubricant Soft rubber sustained
catheters. 4 units fire proof
blanket, 2 units pillow 4 units bed sheet 21.- Paint scheme Inside Primer first
coat and two coats of White fireproof painting 22 A.-) Accessories a) ONE (1) HP Compressor as described. Ingersoll
Rand Series 6R High Pressure Air- cooled Air compressor according and
improving specifications listed . b) Working pressure:
up to 413.7 bar (capable to work to 330-350 bar) c) Motor power : 80 Kw
d) Motor consumption:
74.6 Kw (nearly 75 Kw) e) Output : 4,190
liters per minute (better than 3,900 lit/min) f) Imput supply: 415 v
AC allowance ±10% 50 Hz g) 50 meter Power
supply cable will be provided. h) Operation and
maintenance manual set one set each will be provided j) All valves, gauges,
pipes, adapters, etc. for connecting compressors to HP air bank will be
provided. k) Spares for at least
five years to be provided. 2.- Air reserve cylinder primary air supply
system must have 28 units of 50 litre volume each one @ 360 bar with gauges,
pipes valves and regulators to reduce HP air up to 16 bar will be provided 3.- Oxigen reserve :
20 cylinder 50 liter @ 250 bar will be provided. 22.b.- Miscellaneous (1) Stop watch 06 digit. (2) Monitoring system a)
a) Audio and video recording and monitoring
system will be provided. b)
b) Two in no 14 inch colour monitor will be
provided c)
c) Air protector 10 in no for chamber internal
use will be provided. 23.- Spares:
IBERCO will provide all those spares listed in this paragraph 24.- Guarantee. IBERCO guarantee that
this model will remain in fabrication during 10 years, at least. IBERCO guarantee the
supply of spares for at least 15 years with reasonable price. Yearly increase
of price will not be more than 5%. 25 39 Items
O.K. with all conditions. |
|||||||||||||||||||
|
|
We'd be very glad to attend you at...
|
S.A. Ibιrica de Suministros y Construcciones IBERCO Lotos,
15 30205 -
CARTAGENA (SPAIN)
Phones
+34 968 527070
Fax. + 34 968 501721 |